二氧化硫(SO₂)气体浓度分析仪
二氧化硫(SO₂)气体浓度分析仪
无机气体 AMC 监测
无机气体 AMC 监测
SI5450 将高灵敏度激光光谱技术集成于一套结构紧凑、易于操作且高度可靠的系统中。该分析仪采用中红外激光器,可高精度测量二氧化硫(SO₂)浓度,性能稳定。与传统 AMC 测量技术(如离子迁移谱 IMS、离子色谱、质谱和紫外荧光)相比,SI5450 在响应速度、检测精度和运行成本方面均具有显著优势。传统方法普遍存在响应慢、性能不足、运行成本高的问题。
主要优势
实时监测洁净室、FOUP 及晶圆厂设备中的 AMC
秒级快速、连续分析
几乎无停机时间,耗材成本极低
不受常见晶圆厂化学品干扰
极高测量准确度,适用于事件确认
Gas Detected | Sulfur Dioxide (SO2) |
---|---|
Precision (1σ) | ≤360 ppt (10 sec), ≤120 ppt (100 sec), ≤70 ppt (300 sec) |
Lower Detectable Limit (3σ) | 350 ppt (100 sec), 200 ppt (300 sec), 140 ppt (600 sec) |
Linearity (per IEC 61207) | ±1% |
Accuracy at span | ±5% @ full scale |
Accuracy at zero | ±300 ppt |
Instrument-to-Instrument Consistency | ±5% @ full scale ±300 ppt @ zero |
Measurement Range | 0-1000 ppb |
Measurement Interval* | < 5 seconds |
Sample Flow Rate | ~0.2 slm |
Combined Response Times (T90/10 + T10/90) @ 20 ppb (Humid) | <30 seconds |
Fall Times T90/10 @ 20 ppb (Humid) | <15 seconds |
Measurement Technique | Cavity Ring-Down Spectroscopy |
---|---|
Calibration Period Recommendation | Calibration not required - Initial validation at 6 months, then every 12 months thereafter |
Time Required to Perform Validation | Estimated at <15 minutes per manufacturer's instructions |
Measurement Cell Temperature Control | ±0.005°C |
Measurement Cell Pressure Control | ±0.0002 atm |
Operating Temperature | 15 to 35°C (operating); -10 to 50°C (storage) |
Ambient Humidity | <99% RH non-condensing |
Accessories | Pump (external, included), keyboard (included), mouse (included), LCD monitor (optional), Maintenance Kit (optional) |
Communication Interfaces | RS-232, Ethernet, USB, Modbus, 4-20mA (optional) |
Sample Inlet Connection | 1/4" Stainless Steel Swagelok® Tube Fitting (recommended 1/4" OD PFA Tubing) |
Dimensions | Analyzer: 16.63" w x 8.75" h x 30.25" d (42.24 x 22.22 x 76.84 cm) |
Weight | 75 lbs. (34 kg) including external pump |
Power Requirements | 100–240 VAC, 47–63 Hz (auto-sensing), <400 W (total): 250 W (analyzer), 150 W (pump) at steady state |
Warranty | 12 months |
Certifications | CE conformance |
Country of Manufacture | USA |
Components/Parts | Actions | Frequency |
---|---|---|
Particulate filter | Replace | Every 6 months |
Measurement Validation | Proxy validation | Every 12 months |
Pump | Rebuild (diaphragm) | Preventive maintenance every 2 years |
Fan assemblies | Replace | Preventive maintenance every 2 years |
External Pump | Replace | Every 5 years |